CO2 Analyzers 3100 & 3150
Instruction 19-9339 40
Event Data Recorded
Successful User Zero CO
2
Type (N
2
or Air) and Readings Before and After
Successful User Span CO
2
Target Value, Readings Before and After
Successful User Span O
2
Target Value, Readings Before and After
Failed User Zero CO
2
Type (N
2
or Air) and Reading
Failed User Span CO
2
Target Value, Gas Reading
Failed User Span O
2
Target Value, Gas Reading
Return to Factory Settings
NOTE: If the calibration failed, then try again using a longer purge
time or different target gas. This event log can only be
downloaded and viewed via the optional software. It cannot
be viewed on the analyzer screen.