Instruction 19-9208 3
N
2
O Monitor 3010
2.0 OPERATION
2.1 Important Note
Always ensure that the monitor’s gas inlet connector (Figure 1, Item B)
and outlet (Figure 1, Item D) are unobstructed and open to the atmo-
sphere. Be careful not to breath directly on the monitor while taking a
measurement; otherwise, inaccurate N
2
O readings will result.
2.2 Switching the Monitor ON/OFF
Switch ON the monitor by momentarily pressing the button. Switch the
monitor OFF by pressing the button for at least 3 seconds, or until the
display goes blank. When first switched ON, there is a 30 minute warm-
up period before the N
2
O level is displayed.
Figure 1. Components of the 3010 N
2
O Monitor
A Digital display
B Gas inlet
C Restrictor
D Gas outlet
E TWA alarm light
F IrDA link
G Pump ON light
H ON/OFF button
I Battery charging socket
J Toggle display between
TWA and ppm (f1)
K Calibration / Leak detec-
tion mode (f2)
L Pump ON/OFF (f3)
M – Particulate Filter
2
3
N
2
O M o n i t o r 3 0 1 0
A
L
K
J
B
F
G
H
I
E
C
D
M