CHAPTER 6. CALIBRATION
IMPORTANT: Before performing any calibration procedure ensure
that new batteries are installed or use the optional AC power
adapter. Also ensure that the analyzer is at room temperature and
will be sampling fresh air when turned ON.
6.1. B-Smart Sensors
The PCA
®
3 uses Bacharach’s B-Smart
®
Sensor technology for CO
LOW
, CO
HIGH
,
NO, NO
2
, and SO
2
. The B-Smart
®
Sensor is marked with a 10- or 14-digit
calibration code that can be entered in the instrument’s calibration screen via the
keypad, or with the Fyrite
®
User Software.
Benefits of the B-Smart
®
Sensors include the following:
•
New sensors can be installed without applying gas for calibration.
•
Sensors can be pre-calibrated and installed when needed
•
Sensors can be moved from one analyzer to another.
•
The analyzer’s diagnostics screen shows what sensors are installed,
and their current end-of-life condition (Good, Low, Bad).
B-Smart
®
sensors should be calibrated by an authorized Bacharach Service
Center (see Service Centers on page 99) every 6 months to 1 year to assure that
the analyzer continues to meet its published accuracy specifications. B-Smart
®
sensors, however, can be calibrated in the field if your facility has the necessary
equipment and qualified personnel to perform the procedures described in the
following sections.
6.2. Starting a Calibration
Start any calibration procedure by doing the following.
1. Place the probe in an area of fresh air, turn ON the analyzer, allowing it
to cycle through its 60 second warm-up period. During warm-up, the
analyzer’s operation is checked and the sensors are set to the following
ambient conditions:
•
Oxygen sensor spanned to 20.9%
•
All gas sensors are zeroed
•
The pressure sensor is zeroed
2. Any errors detected during warm-up will be listed on the display
immediately following warm-up. Correct any errors before proceeding.
Refer to page 93 (Error Messages) for a listing of error messages and
their meaning.
3. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
4. Use the ▲▼ buttons to highlight CALIBRATION, and then press ENT to
display the CALIBRATION PASSWORD screen.
Vedere la pagina 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 103 104

Commenti su questo manuale

Nessun commento

Supermicro Placas base manuales de usuario

Empezando en H

Haga clic en el alfabeto a continuación para desplazarse a la lista completa de modelos que comienzan con esa letra

Modelos Tipo de documento
H8DA8
Ficha de datos   Supermicro H8DA8 AMD Dual Opteron 200 Dual U320 Motherboard, 78 paginas
H8DAE-2
Manual de usuario   Supermicro H8DAE-2 motherboard, 82 paginas
H8DAI-2
Manual de usuario   Supermicro H8DA3-2 motherboard, 84 paginas
H8DC8
Manual de usuario   Supermicro H8DCI motherboard, 78 paginas
H8DCR-I
Manual de usuario   Supermicro MBD-H8DCR-I-O motherboard, 78 paginas
H8DME-2
Manual de usuario   Supermicro H8DM8-2 motherboard, 78 paginas
H8DMU+
Manual de usuario   Supermicro H8DMU+ motherboard, 80 paginas
H8DSR-8
Manual de usuario   Supermicro H8DSR-8 motherboard, 74 paginas
H8SGL-F
Manual de usuario   Supermicro H8SGL-F, 74 paginas
H8SSL-I
Manual de usuario   Supermicro H8SSL-I motherboard, 68 paginas