ECA 450 Maintenance
Instruction 0024-9400
61
4 Maintenance
Field maintenance of the ECA 450 is limited to the following:
• Battery charging
• Water Trap Filter replacement
• Printer Paper replacement
• Sensor replacement
• Cleaning the probe
• Resetting the microprocessor
• Replacing the internal particulate filter
All other maintenance should be performed by an authorized Bacharach
Service Center (Section 6.3).
4.1 Battery Charging
A fully charged battery pack will power the analyzer for a minimum of
8 hours. When the battery pack is just about depleted, the message “LOW
BATTERY” will appear at the bottom of the screen; after which, the analyzer
will operate from 5 to 15 minutes before automatically shutting itself off.
Recharge the analyzer’s internal battery pack by first plugging the supplied
AC adapter in a source of 100–240 VAC, 50/60 Hz power. A green LED on the
adapter should glow, indicating that it is operational. Then insert the
adapter’s output plug into the analyzer’s POWER connector (Figure 1-1).
A fully discharged battery pack should take approximately 11 hours to
recharge. Note that a protection circuit within the analyzer prevents the
battery pack from being over charged.
If the battery pack becomes depleted while the analyzer is being used, or if a
continuous run time of over 8 hours is required, then the AC adapter can be
used to both run the analyzer and charge the battery pack at the same time.